Página 2

iCPlate22•General1022.4 Packaging and transport Always forward iCPlate2 in the original device case to avoid damage. Safeguard the measuring foo

Página 3

iCPlate22•GeneralEN112.6 Accessories To perform and document quality control, it is often necessary to save the measured data on a PC.2.6.1 Pl

Página 5

iCPlate23•OperationEN133 OPERATION 13 3.1 FUNCTIONAL ELEMENTS 15 3.2 FIRST STEPS FOR USE 16 3.2.1 Power-saving mode 16 3.2.2 Toolbar

Página 7 - 2 General

iCPlate2EN153 Operation 3.1 Functional elements 3iCPlate23•OperationDisplayPositioning Aid<ENTER> button<DOWN> button<RESET> but

Página 8 - 2•General

3.2 First steps for use Release the measuring head by sliding the locking device on the left side forward. After opening initially or after pr

Página 9

iCPlate23•OperationEN173.2.2 Toolbar The toolbar appears on the right on the display and displays the functions available using symbols. A curs

Página 10 - 2.4 Packaging and transport

iCPlate23•Operation183 3.2.3.2 Move the sub-cursorShift the sub-cursor from left to rightShift the sub-cursor from right to leftShift the sub-cursor

Página 11 - 2.6 Accessories

iCPlate23•OperationEN193 3.2.3.3 Device SettingsStandard platePolyester platePaperFilmScreen ruling in lines/cmScreen ruling in lines/inchRegular scr

Página 13 - Appendix

iCPlate3•Operation203 3.2.3.4 Other functionsTransfer current record (binary image or characteristic curve) to the host PCDevice is calculating (duri

Página 14

iCPlate2EN213.2.3.6 Status informationStandard platePolyester platePaperFilmPositive dot %Negative dot %Regular screeningStochastic screeningRed illum

Página 15 - 3 Operation

iCPlate22233.3 Adjusting device settings After unpacking or after pressing the <RESET> button, you can use iCPlate2 to begin measuring stan

Página 16 - 3.2 First steps for use

iCPlate2EN233.4 FOGRA Measuring Bar FMB setting Select the measurement mode in the device setting (chapter 3.2.3.3): iCPlate2 measurement mode

Página 17 - 3.2.2 Toolbar

iCPlate2243.5 Inputting a reference curve The user can switch to ‘Input a reference curve’ by selecting the function in the settings window (se

Página 18 - 3•Operation

iCPlate2EN25Important:In normal circumstances the reference value is set to be equal to the nomi-nal value to obtain a linear plate copy. However, for

Página 19

iCPlate2263.6 Measuring3.6.1 General Position the instrument by means of the positioning target on the patch and lower the measuring head. The

Página 20 - 3.2.3.4 Other functions

iCPlate2EN273iCPlate23•OperationImportant:In a regular screen, the screen ruling, dot size and screen angle will always be shown if the dot area has c

Página 21

iCPlate2283.6.2 Calibration As the iCPlate2 calibration is done automatically, no calibration or “zeroing” on the plate is needed. After inputti

Página 22

iCPlate2EN293iCPlate23•Operation3.6.4 Measuring polyester plates Standard plates and polyester plates are measured with a red LED. Use of the bl

Página 23

iCPlate21•Table of contentsEN31234Table of Contents General OperationAppendix1 TABLE OF CONTENTS 32 G

Página 24

iCPlate2303.6.6 Measuring paper For paper measurements, automatic color selection allows for quicker work, as no switching between the CMY color

Página 25 - 30, 60 and 90

iCPlate2EN31iCPlate23•Operation3 3.6.7.3 Dot diameter The diameter of the dots is displayed in µm. The calculation is based on a circular dot of

Página 26 - 3.6 Measuring

iCPlate2323.7 Visually checking the dot After a measurement the picture will be held in the device’s memory until the next measurement and can b

Página 27

iCPlate2EN33iCPlate23•Operation33.8 Measuring a plate characteristic curve The user can change from the standard display to the plate characteris

Página 28

iCPlate234Example: Measuring a plate characteristic curve with the sample points 5, 10, 30, 60 and 90 (45° reference curve): • Select the function

Página 29 - 3.6.5 Measuring film

iCPlate2EN35iCPlate23•Operation33.9 iCPlate2 maintenance and care3.9.1 Reset If the microprocessor becomes blocked (e.g. after changing batter

Página 30 - 3.6.7 Measurement Values

iCPlate236 3.9.2 Replacing the batteries The batteries must typically be replaced after 30,000 measurements. The iCPlate2 monitors the battery

Página 31

iCPlate2EN37iCPlate23•Operation4 APPENDIX 37 4.1 TECHNICAL DATA 39-40 4.2 SERIAL INTERFACE 41 4.3 UPGRADE TO ICPLATE2 XT 42 4.4 DECLARATION

Página 32

iCPlate238

Página 33

iCPlate2EN39iCPlate24•Appendix4 Appendix4.1 Technical data iCPlate2 X iCPlate2 XTFunctions Dot area % X X Screen ruling in lines/cm or l

Página 34

iCPlate21•Table of contents4 3.7 VISUALLY CHECKING THE DOT 31 3.8 MEASURING A PLATE CHARACTERISTIC CURVE 32-33 3.9 iCPlate2 MAINTENANCE AND CARE

Página 35 - 3.9.1 Reset

iCPlate2404iCPlate24•Appendix iCPlate2 X iCPlate2 XT Dot size range (FM) 10 µm – 50 µm Dot diameter resolution 1 µm Screen angle resoluti

Página 36 - 3.9.2 Replacing the batteries

iCPlate2EN414.2 Serial interface The iCPlate2 is equipped with a RS232 interface. To obtain access to the serial interface socket, remove the

Página 37

iCPlate2424.3 Upgrade to iCPlate2 XT The iCPlate2 X may be upgraded to an iCPlate2 XT. The functions which you did not purchase initially may

Página 38

iCPlate2EN43iCPlate24•Appendix4.4 Declaration of conformityCE-DECLARATION OF CONFORMITYThe undersigned, representing the following manufacturer: X-

Página 39 - 4 Appendix

iCPlate2444.5 Warranty Information X-Rite, Incorporated (“X-Rite”) warrants each instrument manufactured to be free of defects in material and w

Página 40 - 4•Appendix

iCPlate22•GeneralEN51234Table of Contents General OperationAppendix2 GENERAL 5 2.1 INTRODUCTION 7 2.2 BRIEF GUIDE TO THE MANUAL 8 2.3 SAFETY

Página 42 - 4.3 Upgrade to iCPlate2 XT

iCPlate22•GeneralEN72 General 2.1 IntroductionCongratulations! You have just acquired the portable plate

Página 43

iCPlate22•General82.2 Brief guide to the manual Descriptions marked with a only apply to the iCPlate2 XT and are not part of the iCPlate2 X

Página 44

iCPlate22•GeneralEN922.3 Safety2.3.1 Warning For safety reasons it is absolutely necessary to read through the user’s guide and all of the

Comentários a estes Manuais

Sem comentários